Skip to content
All Treatments
Laser

BLESSING Diagonal-Needle RF & Injection

Korean platform combining diagonal-needle RF, direct injection, and subcision-like action

Duration
45-75 min (表面麻酔を含む。機器処置は約10-15 min)
Downtime
2-7 days
Sessions
3 sessions前後 (4-6 wk intervals、適応・併用製剤で調整)

About This Treatment

BLESSING is a microneedle RF platform developed by South Korean manufacturer Cellah Medical. It includes DRS with diagonally entering needles, VRS with vertical needles, and ORS with a single-pin electrode. The manufacturer describes DRS as combining RF delivery, injectable placement, and a subcision-like scar-release effect.

The system received U.S. FDA 510(k) clearance in May 2025, but only for electrocoagulation and hemostasis in dermatologic and general surgical procedures. The clearance does not establish efficacy for cosmetic rejuvenation, acne-scar subcision, or drug delivery. No BLESSING-specific peer-reviewed human clinical study was identified as of August 17, 2026; counseling must distinguish indirect RF-microneedling and conventional-subcision evidence from manufacturer claims.

Mechanism of Action

FDA K243176 describes bipolar or monopolar RF at 1 or 2 MHz, with tissue resistance producing heat and focal coagulation. The 0.28-mm D-N10 DRS tip is designed to enter the skin diagonally. Cellah says this track mechanically crosses shallow fibrotic tethering while a lumen delivers an injectate together with RF, but incremental drug-delivery and scar-release benefits have not been validated in product-specific peer-reviewed human trials and are not part of the FDA-cleared indication.

Indications

AcnePoresWrinkles / fine linesSagging / contourGeneral indication (see detailed description)

Expected Results

No BLESSING-specific response rate or durability has been established in peer-reviewed human studies. A 2025 systematic review of 16 fractional RF-microneedling studies and 481 patients found the device class likely effective for acne scars, but studies were heterogeneous and further randomized trials are needed. Expected changes should therefore be described as class-inferred, gradual improvement in scars, texture, pores, and firmness over several weeks to 1-3 months after a series. Incremental benefit over devices such as Potenza, conventional subcision, or injection alone is unknown.

Clinical Evidence

No product-specific peer-reviewed evidence
Treatment database evidence review (2026)
No BLESSING-specific peer-reviewed human clinical evidence identified. PubMed evidence search
Clinical improvement was reported in this study (see original paper for details).
Regulatory / official
U.S. Food and Drug Administration (2025)
510(k) Summary K243176: BLESSING System. FDA 510(k) Premarket Notification
Clinical improvement was reported in this study (see original paper for details).
Regulatory / official
U.S. Food and Drug Administration (2025)
Potential Risks with Certain Uses of Radiofrequency (RF) Microneedling. FDA Safety Communication
Clinical improvement was reported in this study (see original paper for details).
Indirect / device class
Niaz G, Ajeebi Y, Alshamrani HM, Khalmurad M, Lee K (2025)
Fractional Radiofrequency Microneedling as a Monotherapy in Acne Scar Management: A Systematic Review of Current Evidence. Clin Cosmet Investig Dermatol
Clinical improvement was reported in this study (see original paper for details).
Indirect / device class
Ahramiyanpour N, Rastaghi F, Parvar SY, Sisakht AK, Hosseini SA, Amani M (2023)
Subcision in acne scarring: A review of clinical trials. J Cosmet Dermatol
Clinical improvement was reported in this study (see original paper for details).
Manufacturer material
Cellah Medical / KOTRA BuyKorea (2026)
BLESSING: Diagonal Needle RF & Injection System. Manufacturer product listing
Clinical improvement was reported in this study (see original paper for details).
Indirect / device class
GMO Beauty / Kireipass byGMO (2026)
2025年美容医療人気ランキングおよび2026年美容医療トレンド予測. Commercial platform trend report
Clinical improvement was reported in this study (see original paper for details).

Risks & Side Effects

Erythema, edema, warmth, pain, pinpoint bleeding, bruising, crusting, pigment change, burns, blistering, infection, herpes reactivation, scarring, fat loss, contour change, numbness, and nerve injury may occur. Diagonal-needle or subcision-like manipulation adds potential vessel, nerve, and deeper-tissue injury. Injectates add product-specific risks including allergy, delayed inflammation, nodules or granuloma, infection, intravascular injection or occlusion, and necrosis; sterility, local authorization, injection plane, and emergency readiness must be confirmed.